abberior MIRAVA POLYSCOPE 3D STED Confocal Superresolution Microscope
ICOB Imaging Core | Instrument Overview, Sample Preparation, Reservation Policies, and Fees
Document status: Pilot Operation Draft
The reservation policies and fees described on this page apply during the pilot operation period. Any changes will be subject to announcements by the instrument administrators.
1. Instrument Overview
The abberior MIRAVA POLYSCOPE, developed by abberior instruments in Germany, integrates confocal microscopy with stimulated emission depletion (STED) microscopy. The system supports Confocal, 2D STED, and 3D STED imaging, enabling improved lateral (XY) and axial (Z) spatial resolution. It is suitable for imaging cytoskeletal structures, organelles, protein distributions, and other fine structures in cells and tissues.
1.1 Resolution Reference
- 2D STED: Lateral (XY) resolution of approximately 20–30 nm.
- 3D STED: Three-dimensional (XYZ) resolution of approximately 60–90 nm; isotropic resolution of approximately 70 nm may be achieved under ideal conditions.
2. System Configuration and Advanced Features
| Item | System Configuration |
|---|---|
| Instrument model | abberior MIRAVA POLYSCOPE |
| Microscope platform | Olympus IX83 inverted microscope |
| Imaging modes | Confocal, 2D STED, 3D STED |
| Excitation lasers | 405 nm (continuous wave, CW); 485, 560, and 640 nm (pulsed) |
| STED depletion laser | 775 nm |
| Detectors | Three APD-based spectral detectors (avalanche photodiode detectors) |
| Objective lenses | Olympus UPLXAPO 60×/1.42 NA, WD 0.15 mm Olympus UPLXAPO 60×/1.20 NA, WD 0.28 mm |
| Advanced features | FLEXPOSURE adaptive illumination, RAYSHAPE adaptive optics, TRUESHARP deconvolution |
| Acquisition software | abberior LiGHTBOX |
2.1 Advanced Imaging Technologies
- FLEXPOSURE (adaptive illumination): Adjusts illumination according to image signals and structures to reduce unnecessary laser exposure and photobleaching.
- RAYSHAPE (adaptive optics): Compensates for optical aberrations and helps improve imaging quality in thick specimens and during 3D STED acquisition.
- TRUESHARP (deconvolution): Provides image deconvolution and detail enhancement. For quantitative image analysis, users are advised to retain the original images and processing parameters.
3. Recommendations for STED Sample Preparation
The system is not currently equipped with environmental controls required for live-cell incubation. Therefore, fixed samples are recommended at this stage. Before preparing STED samples, users are encouraged to consult the instrument administrators regarding fluorophore selection, objective lenses, sample thickness, and mounting conditions.
3.1 Fluorophores and Excitation Wavelengths
The following fluorophores are recommended for initial consideration. All three abberior STAR dyes listed below are compatible with the system's 775 nm STED depletion laser.
| Fluorophore | Excitation Wavelength | Imaging Compatibility |
|---|---|---|
| abberior STAR RED | 640 nm | Confocal and STED |
| abberior STAR ORANGE | 560 nm | Confocal and STED |
| abberior STAR 460L | 485 nm | Confocal and STED |
| DAPI / Hoechst | 405 nm | Confocal only; not suitable for STED imaging with this system's 775 nm depletion laser |
3.2 Mounting Media and Coverslips
| Item | Recommendations and Notes |
|---|---|
| Mounting medium | abberior MOUNT SOLID ANTIFADE (priority manufacturer-recommended option; requires time to cure). abberior MOUNT LIQUID ANTIFADE (non-hardening; manufacturer-recommended option; seal around the coverslip edges with nail polish or an appropriate sealant). |
| Coverslips | No. 1.5H high-precision glass coverslips are recommended (thickness 170 µm). |
4. Operating Modes and Fees
Access to the instrument is divided into the following three categories according to user certification, operating arrangements, and time of use.
| Operating Mode | Eligibility / Time | Reservation Method | Fee |
|---|---|---|---|
| Independent operation by certified users: Regular working hours | Certified users from across Academia Sinica; 08:00–17:00 | IIS Instrument Reservation System, under ICOB: Abberior Mirava Polyscope (3D STED Confocal) |
NT$500/hour |
| Independent operation by certified users: Outside regular working hours | Only certified users affiliated with the Institute of Cellular and Organismic Biology (ICOB) | IIS Instrument Reservation System, under ICOB: Abberior Mirava Polyscope (Non-working time) |
NT$250/hour |
| Administrator-assisted operation / User training | Users without certification or users requiring assistance; advance coordination required | Contact both instrument administrators by email | NT$1,000/hour |
5. Reservation Schedule and Priority Allocations During Regular Working Hours
5.1 Reservation Policy: Priority Booking for the Following Week; Open Booking for the Current Week
| Applicant Institute | Eligible Booking Weeks | Eligibility and Booking Rules |
|---|---|---|
| Priority booking units | Current week and following week | Certified users from the priority booking institutes may reserve instrument time for both the current and following week. Reservations for the following week count toward the institute's weekly priority booking quota, while reservations for the current week do not. |
| Other Academia Sinica units | Current week only | Certified users from all other Academia institutes may only book remaining available time slots within the current week. Advance booking for the following week is not permitted. |
5.2 Reservation Schedule by Week
The following table illustrates three consecutive booking weeks. Rows indicate when a reservation is made, while columns indicate the intended week of instrument use.
| Usage Week → Booking Week ↓ |
Week 1 | Week 2 | Week 3 | Week 4 |
|---|---|---|---|---|
| Week 1 | Open to all certified users | Priority booking units | — | — |
| Week 2 | — | Open to all certified users | Priority booking units | — |
| Week 3 | — | — | Open to all certified users | Priority booking units |
5.3 Weekly Priority Booking Allocations
During the preceding week, each priority booking unit may reserve time slots during regular working hours for the following week, up to its allocated weekly limit.
| Priority Booking Unit | Priority Allocation (Hours/Week) |
|---|---|
| Institute of Cellular and Organismic Biology (ICOB) | 6 |
| Institute of Molecular Biology (IMB) | 3 |
| Institute of Biomedical Sciences (IBMS) | 3 |
| Institute of Plant and Microbial Biology (IPMB) | 3 |
| Institute of Biological Chemistry (IBC) | 2 |
| Genomics Research Center (GRC) | 2 |
| Total | 19 |
5.4 Per-Laboratory Booking Limit and Reservation Review
- Each laboratory may reserve a maximum of 4 hours per week during regular working hours, including both advance priority bookings and current-week bookings.
- Reservations must be made using the laboratory's IIS account, rather than an individual IIS account, so that bookings can be verified by laboratory and affiliated unit.
- Administrators will verify user eligibility and booked hours against the reservation records. Reservations that do not comply with these policies may require modification or may be rejected.
5.5 Use Outside Regular Working Hours
- Reservations and independent operation outside regular working hours are restricted to certified ICOB users.
- Reservation system item:
Abberior Mirava Polyscope (Non-working time). - Fee: NT$250/hour.
- Hours booked outside regular working hours do not count toward the weekly priority allocations for regular working hours.
6. How to Make a Reservation or Request Assistance
6.1 Certified Users: Self-Service Reservations
- Confirm that you have obtained certification to operate this instrument and that you are eligible to reserve the desired time slot.
- Check the intended week of use: reservations for the following week are initially available to priority booking units within their allocated hours, while remaining time slots in the current week are open to all certified Academia Sinica users.
- Go to the Academia Sinica Instrument Reservation System, select Equipment Reservation, and choose the appropriate item:
- Regular working hours:
Abberior Mirava Polyscope (3D STED Confocal) - Outside regular working hours:
Abberior Mirava Polyscope (Non-working time)
- Regular working hours:
- Complete the reservation using your laboratory's IIS account (not your personal IIS account), and comply with the applicable booking limits and reservation review decisions.
6.2 Administrator-Assisted Operation / User Training: Contact in Advance
If you need assistance with instrument operation or wish to arrange user training, please email both instrument administrators to coordinate a suitable time:
- Po-Yen Lin: pylin@as.edu.tw
- Wei-Chen Chu: weichen01@as.edu.tw
- Fee: NT$1,000/hour.
- Scheduling is subject to instrument availability, the reserved hours allocated to priority booking units, and the administrators' availability. When contacting the administrators, please briefly describe your sample type, fluorophores, mounting medium, preferred imaging mode (Confocal / 2D STED / 3D STED), and expected usage time.